DC Volts and DC Current Gain Verification

This procedure verifies DC volts and DC current gain.

  1. Make sure you have read Test Considerations.
  2. Connect the calibrator to the input terminals. For the 34461A/65A/70A, use the front panel input terminals and select the front input terminals with the Front/Rear switch.
  3. Configure each function and range in the order shown in the table below corresponding to the DMM model number. Provide the input shown in the table.
  4. Make a measurement and return the result. Compare measurement results to the test limits shown below. (Be certain to allow for appropriate source settling.)

Values for 34460A DMM
Input Error from Nominal
Voltage Function[1] Range Quick
Check
24 hours 90 days 1 year 2 years
100 mV DC Volts 100 mV ± 10 µV ± 13.5 µV ± 15.5 µV ± 18 µV
-100 mV
1 V 1 V ± 39 µV ± 70 µV ± 90 µV ± 115 µV
-1 V
4 V 10 V ± 140 µV ± 250 µV ± 350 µV ± 450 µV
10 V 10 V Q ± 290 µV ± 550 µV ± 800 µV ± 1.1 mV
-10 V
100 V 100 V ± 3.6 mV ± 7.1 mV ± 9.1 mV ± 11.6 mV
-100 V Q
1000 V 1000 V ± 36 mV ± 75 mV ± 95 mV ± 120 mV
-500 V ± 21 mV ± 42.5 mV ± 52.5 mV ± 62.5 mV

[1]Select 10 PLC.
Q: Quick performance verification test point.

Values for 34460A DMM
Input Error from Nominal
Current Function[1] Range Quick
Check
24 hours 90 days 1 year 2 years
2 A DC Current 3 A ± 4.2 mA ± 4.6 mA ± 4.6 mA ± 5.2 mA
1 A 1 A Q ± 560 µA ± 900 µA ± 1.1 mA ± 1.3 mA
100 mA 100 mA ± 14 µA ± 35 µA ± 55 µA ± 65 µA
10 mA 10 mA Q ± 2.7 µA ± 5 µA ± 7 µA ± 8 µA
1 mA 1 mA ± 0.13 µA ± 0.36 µA ± 0.56 µA ± 0.66 µA
100 µA 100 µA ± 0.03 µA ± 0.065 µA ± 0.075 µA ± 0.085 µA

[1]Select 10 PLC.
Q: Quick performance verification test point.

Values for 34460A DMM
Input Error from Nominal
Resistance Function[1] Range Quick
Check
24 hours 90 days 1 year 2 years
100 Ω 4-wire Ohms 100 Ω ± 10 mΩ ± 18 mΩ ± 21 mΩ ± 24 mΩ
1 kΩ 1 kΩ Q ± 38 mΩ ± 120 mΩ ± 150 mΩ ± 180 mΩ
10 kΩ 10 kΩ ± 350 mΩ ± 1.2 Ω ± 1.5 Ω ± 1.8 Ω
100 kΩ 100 kΩ ± 3.5 Ω ± 12 Ω ± 15 Ω ± 18 Ω
1 MΩ 1 MΩ ± 40 Ω ± 120 Ω ± 150 Ω ± 180 Ω
10 MΩ 10 MΩ Q ± 1.6 kΩ ± 2.1 kΩ ± 4.1 kΩ ± 6.1 kΩ
100 MΩ 2-wire Ohms 100 MΩ ± 310 kΩ ± 810 kΩ

[1]Select 10 PLC.
Q: Quick performance verification test point.

Values for 34461A DMM
Input Error from Nominal
Voltage Function[1] Range Quick
Check
24 hours 90 days 1 year 2 years
100 mV DC Volts 100 mV ± 6 µV ± 7.5 µV ± 8.5 µV ± 10 µV
-100 mV
1 V 1 V ± 26 µV ± 37 µV ± 47 µV ± 62 µV
-1 V
4 V 10 V ± 100 µV ± 130 µV ± 190 µV ± 230 µV
10 V Q ± 190 µV ± 250 µV ± 400 µV ± 500 µV
-10 V
100 V 100 V ± 2.6 mV ± 4.1 mV ± 5.1 mV ± 6.6 mV
-100 V Q
1000 V 1000 V ± 26 mV ± 45 mV ± 55 mV ± 70 mV
-500 V ± 16 mV ± 27.5 mV ± 32.5 mV ± 40 mV

[1]Select 10 PLC.
Q: Quick performance verification test point.

Values for 34461A DMM
Input Error from Nominal
Current Function[1] Range Quick
Check
24 hours 90 days 1 year 2 years
2 A DC Current 3 A ± 4.2 mA ± 4.6 mA ± 4.6 mA ± 5.2 mA
1 A 1 A Q ± 560 µA ± 900 µA ± 1.1 mA ± 1.3 mA
100 mA 100 mA ± 14 µA ± 35 µA ± 55 µA ± 65 µA
10 mA 10 mA Q ± 2.7 µA ± 5 µA ± 7 µA ± 8 µA
1 mA 1 mA ± 0.13 µA ± 0.36 µA ± 0.56 µA ± 0.66 µA
100 µA 100 µA ± 0.03 µA ± 0.065 µA ± 0.075 µA ± 0.085 µA

[1]Select 10 PLC.
Q: Quick performance verification test point.

Values for 34461A DMM
Input Error from Nominal
Resistance Function[1] Range Quick
Check
24 hours 90 days 1 year 2 years
100 Ω 4-wire Ohms 100 Ω ± 6 mΩ ± 12 mΩ ± 14 mΩ ± 16 mΩ
1 kΩ 1 kΩ Q ± 25 mΩ ± 90 mΩ ± 110 mΩ ± 130 mΩ
10 kΩ 10 kΩ ± 250 mΩ ± 900 mΩ ± 1.1 Ω ± 1.3 Ω
100 kΩ 100 kΩ ± 2.5 Ω ± 9 Ω ± 11 Ω ± 13 Ω
1 MΩ 1 MΩ ± 30 Ω ± 90 Ω ± 110 Ω ± 130 Ω
10 MΩ 10 MΩ Q ± 1.6 kΩ ± 2.1 kΩ ± 4.1 kΩ ± 6.1 kΩ
100 MΩ 2-wire Ohms 100 MΩ ± 310 kΩ ± 810 kΩ

[1]Select 10 PLC.
Q: Quick performance verification test point.

 

 

Values for 34465A DMM
Input Error from Nominal
Current Function[1] Range Quick
Check
24 hours 90 days 1 year 2 years
2 A DC Current 3 A ± 4.2 mA ± 4.6 mA ± 4.6 mA ± 5.2 mA
1 A 1 A Q ± 560 µA ± 800 µA ± 900 µA ± 1.1 mA
100 mA 100 mA ± 14 µA ± 35 µA ± 55 µA ± 65 µA
10 mA 10 mA Q ± 2.7 µA ± 5 µA ± 7 µA ± 8 µA
1 mA 1 mA ± 0.1 µA ± 0.35 µA ± 0.55 µA ± 0.65 µA

[1]Select 10 PLC.
Q: Quick performance verification test point.

Values for 34465A DMM
Input Error from Nominal
Voltage Function[1] Range Quick
Check
24 hours 90 days 1 year 2 years
100 mV DC Volts 100 mV ± 6 µV ± 7.5 µV ± 8.5 µV ± 10 µV
-100 mV
1 V 1 V ± 19 µV ± 29 µV ± 39 µV ± 54 µV
-1 V
4 V 10 V ± 70 µV ± 120 µV ± 160 µV ± 220 µV
10 V Q ± 130 µV ± 240 µV ± 340 µV ± 490 µV
-10 V
100 V 100 V ± 2.6 mV ± 4.1 mV ± 4.6 mV ± 6.1 mV
-100 V Q
1000 V 1000 V ± 36 mV ± 51 mV ± 56 mV ± 71 mV
-500 V ± 16 mV ± 23.5 mV ± 26 mV ± 33.5 mV

[1]Select 10 PLC.
Q: Quick performance verification test point.

Values for 34465A DMM
Input Error from Nominal
Resistance Function[1] Range Quick
Check
24 hours 90 days 1 year 2 years
100 Ω 4-wire Ohms 100 Ω ± 6 mΩ ± 9 mΩ ± 10 mΩ ± 11 mΩ
1 kΩ 1 kΩ Q ± 25 mΩ ± 35 mΩ ± 45 mΩ ± 55 mΩ
10 kΩ 10 kΩ ± 250 mΩ ± 350 mΩ ± 450 mΩ ± 550 mΩ
100 kΩ 100 kΩ ± 2.5 Ω ± 3.5 Ω ± 4.5 Ω ± 5.5 Ω
1 MΩ 1 MΩ ± 25 Ω ± 65 Ω ± 75 Ω ± 85 Ω
10 MΩ 10 MΩ ± 1.1 kΩ ± 2.1 kΩ ± 2.6 kΩ ± 3.1 kΩ
10 MΩ 2-wire Ohms 10 MΩ Q
100 MΩ 100 MΩ ± 101 kΩ ± 201 kΩ ± 301 kΩ

± 401 kΩ

[1]Select 10 PLC.
Q: Quick performance verification test point.

Values for 34470A DMM
Input Error from Nominal
Current Function[1] Range Quick
Check
24 hours 90 days 1 year 2 years
2 A DC Current 3 A ± 4.2 mA ± 4.6 mA ± 5.2 mA
1 A 1 A Q ± 560 µA ± 800 µA ± 900 µA ± 1.1 mA
100 mA 100 mA ± 14 µA ± 35 µA ± 55 µA ± 65 µA
10 mA 10 mA Q ± 2.7 µA ± 5 µA ± 7 µA ± 8 µA
1 mA 1 mA ± 0.1 µA ± 0.35 µA ± 0.55 µA ± 0.65 µA

[1]Select 10 PLC.
Q: Quick performance verification test point.

Values for 34470A DMM
Input Error from Nominal
Voltage Function[1] Range Quick
Check
24 hours 90 days 1 year 2 years
100 mV DC Volts 100 mV ± 6 µV ± 7.5 µV ± 8 µV
-100 mV
1 V 1 V ± 14 µV ± 19 µV ± 24 µV ± 29 µV
-1 V
4 V 10 V ± 52 µV ± 72 µV ± 84 µV ± 100 µV
10 V Q ± 100 µV ± 150 µV ± 180 µV ± 220 µV
-10 V
100 V 100 V ± 2.6 mV ± 3.8 mV ± 4.4 mV ± 4.6 mV
-100 V Q
1000 V 1000 V ± 36 mV ± 48 mV ± 54 mV ± 56 mV
-500 V ± 16 mV ± 22 mV ± 25 mV ± 36 mV

[1]Select 10 PLC.
Q: Quick performance verification test point.

Values for 34470A DMM
Input Error from Nominal
Resistance Function[1] Range Quick
Check
24 hours 90 days 1 year 2 years
100 Ω 4-wire Ohms 100 Ω ± 6 mΩ ± 9 mΩ ± 10 mΩ ± 11 mΩ
1 kΩ 1 kΩ Q ± 25 mΩ ± 35 mΩ ± 45 mΩ ± 55 mΩ
10 kΩ 10 kΩ ± 250 mΩ ± 350 mΩ ± 450 mΩ ± 550 mΩ
100 kΩ 100 kΩ ± 2.5 Ω ± 3.5 Ω ± 4.5 Ω ± 5.5 Ω
1 MΩ 1 MΩ ± 25 Ω ± 65 Ω ± 75 Ω ± 85 Ω
10 MΩ 10 MΩ ± 1.1 kΩ ± 2.1 kΩ ± 2.6 kΩ ± 3.1 kΩ
10 MΩ 2-wire Ohms 10 MΩ Q
100 MΩ 100 MΩ ± 101 kΩ ± 201 kΩ ± 301 kΩ ± 401 kΩ

[1]Select 10 PLC.
Q: Quick performance verification test point.